No flow coming from ACQUITY BSM QSM - WKB46463

Article number: 46463

SYMPTOMS

  • No solvent comes out of the BSM when flow is set
  • No solvent comes out of the prime waste line during priming
  • Solvent does not flow out of the top of the check valve when removed from the pump

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC M-Class
  • ACQUITY Arc

CAUSE

Blocked primary check valve cartridge   

FIX or WORKAROUND

  1. Replace the primary check valve cartridge. (WKB275812)
  2. As a temporary fix, you can also sonicate the check valves in 50/50 Methanol/Water (WKB85138)

ADDITIONAL INFORMATION

This same problem has been observed with a faulty GPV. The solution is to replace the faulty check valve (700005161).

Part numbers for primary check valves

  • ACQUITY UPLC / nano with no I2V               700002596
  • ACQUITY UPLC with I2V / ACQUITY H-Class   700003563 
  • ACQUITY H-Class Bio / ACQUITY M-Class       700005414
  • ACQUITY I-Class / THF/Hexane                   700004139
    • ACQUITY QSM-R                                   700010664

For I2V systems, if replacing the check valve does not correct the problem, replace the I2V actuator, see ACQUITY H-Class unstable or no pressure.  Before replacing the i2V actuator, make sure the connection to the personality PCB is secure.  