No pressure or unstable pressure with ACQUITY - WKB50754

Article number: 50754

SYMPTOMS

  • No pressure or unstable pressure after rebuilding the pumps and replacing the check valve cartridge

ENVIRONMENT

  • H-Class QSM
  • I-Class BSM
  • ACQUITY BSM

CAUSE

Leaking primary i2V assembly

FIX or WORKAROUND

1. Make sure i2V check valve is properly installed 

2. Replace the i2V check valve assembly.

H-Class: 700005162

H-Class Bio: 700005412

I-Class and ACQUITY Classic: 700003557

ADDITIONAL INFORMATION

Before replacing the i2V actuator, make sure the connection to the personality PCB is secure.

Not able to find a solution? Click here to request help.