"Charge Amplifier resident memory Device not Connected" error with ACQUITY UPLC PDA - WKB46466

Article number: 46466

SYMPTOMS

  • "Charge Amplifier resident memory Device not Connected" error
  • "Lamp lighting failure" error

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC PDA eλ Detector
  • 2998 Photodiode Array Detector

CAUSE

The lamp is bad.

FIX or WORKAROUND

Replace the lamp.

ADDITIONAL INFORMATION

High Brightness Deuterium Lamp for PDA is part 700009330, or order PM kit (201000281).

High Brightness Deuterium Lamp (with ID chip) for thermally managed PDA is part 700005269, or order PM kit (201000273).