ACQUITY Sample Organizer communication failure - WKB46454
Article number: 46454
SYMPTOMS
- ACQUITY UPLC Sample Organizer communication failure error message
- Resetting the instrument does not resolve the issue
- Reconfiguring instrument in the console does not resolve the issue
- Rebooting the system and the LAC/E32 does not resolve the issue
ENVIRONMENT
- ACQUITY UPLC Common Platform Sample Organizer - CPSO
- ACQUITY UPLC Sample Organizer - SO
- ACQUITY UPLC H-Class Sample Manager - FTN
- ACQUITY UPLC Sample Manager
CAUSE
Communication issue between the Sample Organizer and the Sample Manager.
FIX or WORKAROUND
- Power-off the Sample Organizer.
- In the ACQUITY Console, click Sample Manager > Configure > Sample Organizer. Set it to "none".
- Reset the Sample Manager.
- Power-on the Sample Organizer. Wait a couple of minutes.
- In the ACQUITY Console, click Sample Manager > Configure > Sample Organizer. Choose the Sample Organizer serial number from the drop-down menu.
- Once the Sample Organizer is reconfigured in the console, it will communicate successfully with the system.
ADDITIONAL INFORMATION
id46454, A-30SM, comms, UPCPORG, UPISMFTN, UPORGANIZE, UPORGIVD, UPPFTN, UPSMFTN, UPSMFTN+