Skip to main content
Waters

How to replace check valve on QSM/ BSM - steps and video - WKB275812

Article number: 275812

OBJECTIVE or GOAL

How to replace check valve on QSM/ BSM

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC M-Class
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

PROCEDURE

NOTE: Each check valve (CV) has a PEEK washer that needs to be inserted (flat side down) onto the top of the CV prior to installation. 

acc.JPGI2v.JPG

  1. Power off the QSM/BSM
  2. For primary pump head rebuilds, raise the solvent lines above the mobile phase reservoir level to prevent siphoning. 
  3. Use 1/4 inch spanner to remove 3 fittings under the check valves
  4. Remove finger tight fitting from active CV housing. A small drop of solvent may leak when this is removed. 
  5. Use 1/4 inch spanner to loosen the active check valve housing and remove it from the pump head. 
  6. Using a T8 torx driver, loosen the 4 screws on top of the active CV housing. 
  7. Remove the old check valve by pulling it upwards.
  8. Place new CV into the CV housing. Ensure the washer is sitting on top of the new CV.
  9. Push in the tabs either side of the CV housing and tighten the 4 screws using the T8 torx drive.
  10. Reinstall the active CV housing into the pump head. 
  11. Using the 1/4 inch spanner, tighten the active CV housing into the pump head with a 1/4 turn. 
  12. Reinstall the finger tight tiffing into the active CV housing. 
  13. Using the 1/2 inch spanner, remove the accumulator CV from the pump head.
  14. Install a new accumulator CV, ensuring the washer is installed on top of the CV.
  15. Using a 1/2 inch spanner, tighten the CV into place.
  16. Reinstall tubing connecting the accumulator CV to the primary transducer. 
  17. Use a 1/4 inch spanner to tighten the fittings the last 1/4 turn.
  18. Reconnect the tubing to the accumulator transducer.
  19. Power on the QSM/BSM.
  20. Prime with MeOH** in one line for at least 5 minutes.
  21. Start the flow in MeOH** to energize check valves, keep instrument under flow for at least 5 minutes.
  22. Return proper phases for analysis.

ADDITIONAL INFORMATION

** Titanium filters should not be used in 100% methanol, a minimum of 5% water is required for Bio systems with Titanium solvent filters.  

When using titanium mobile phase filters (sinkers), Waters recommends methanol with no less than 5% of water to reduce the possibility of titanium solubilization. Or when using methanol as a mobile phase we recommend using stainless steel solvent filters instead of titanium solvent filters.

What are the part numbers for the ACQUITY UPLC H-Class Bio and ACQUITY Premier solvent manager check valves? - WKB261170

Video Replacing the Check Valve

Not able to find a solution? Click here to request help.