SYMPTOMS

  • Pressure is unstable
  • Dynamic leak test pass
  • Sonicating the check valves does not resolve issue
  • Primary pressure trace displays little to no pressure difference on each stroke (see picture)
    clipboard_e42fb092b5b4daa2df213e2dddafbc0b8.png

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC M-Class
  • ACQUITY Premier

CAUSE

Faulty primary i2v check valve

FIX or WORKAROUND

  1. Replace the check valve. Click HERE for detailed steps and a video showing check valve replacement. 
     