Unstable pressure on ACQUITY UPLC - WKB22900
Article number: 22900
SYMPTOMS
- Pressure is unstable
- Dynamic leak test pass
- Sonicating the check valves does not resolve issue
- Primary pressure trace displays little to no pressure difference on each stroke (see picture)

ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY Quaternary Solvent Manager (QSM)
- ACQUITY UPLC M-Class
- ACQUITY Premier
CAUSE
Faulty primary i2v check valve
FIX or WORKAROUND
- Replace the check valve. Click HERE for detailed steps and a video showing check valve replacement.

ADDITIONAL INFORMATION
Part Numbers:
For H-Class: 700005165, i2 Valve Cartridge
For H-Class Bio: 700005414, Cartridge, i2V, BioACQUITY
For M-Class: 700005414, Cartridge, i2V, BioACQUITY
For more options see Graphical Parts Locator