ACQUITY UPLC H-Class CH-30 temperature does not show in Sample Manager control panel - WKB228917

Article number: 228917

SYMPTOMS

  • CH-30 temperature does not show in Sample Manager control panel

ENVIRONMENT

  • ACQUITY H-Class Sample Manager

CAUSE

The system was not rebooted after a change was made.

FIX or WORKAROUND

Reboot the system. Ensure that the CH-30 is powered-on before the Sample Manager.

ADDITIONAL INFORMATION

 

id228917, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPBINARY