ACQUITY SM-FTN error: Sample Fluidics High Pressure Limit (value) - WKB19999
Article number: 19999
SYMPTOMS
- Error: "Sample Fluidics High Pressure Limit (value)"
- High pressure with inject valve in Inject position
- Pressure decreases with inject valve in Load position
- Error "Pump H/W overpressure" occurs following needle seal readiness test
- Error: "Sample Fluidics High Pressure Limit (value)" on ACQUITY Sample Managers does not resolve the issue
- ACQUITY H-Class System Overpressure Error does not resolve the issue
- While performing Characterize Needle Seal the resulting value for the seal force constant is too high (19 N/mm)
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- SM-FTN
- ACQUITY UPLC I-Class
- ACQUITY APC
CAUSE
Clogged seal extension tube (seat port)
Partially clogged Vespel Seat with Anti-Rot (# 405011492)
FIX or WORKAROUND
Replace seal extension tube, 700005234. (H-Class SM-FTN)
Replace seal extension tube, 700006056 (I-Class SM-FTN and ACQUITY APC)
Replace Vespel Seat with Anti-Rot (# 405011492)
ADDITIONAL INFORMATION
Contact Waters for a local service representative to replace this part.
Values for the seal force constant should be between 8-11 N/mm
id19999, A-30SM, UPASMFTN, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO