Baseline spike and pressure drop with ACQUITY Arc - WKB20009
Article number: 20009
SYMPTOMS
- Spike observed in baseline
- Pressure drop observed at same retention time as spike
ENVIRONMENT
- ACQUITY Arc QSM-R
- Arc HPLC
CAUSE
The check valves are sticking.
ADDITIONAL INFORMATION
Other options:
Check Valve, Primary, THF (# 700011074)
Check Valve, Double Ball and Seat, THF (# 700005273)
Check Valve, Primary, Bio (# 700010642)
Check Valve, Double Ball and Seat, Bio (# 700005415)