ACQUITY H-Class System Overpressure Error - WKB16398
Article number: 16398
SYMPTOMS
- ACQUITY H-Class System Overpressure Error with a low flow rate
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
The Seat Port Assembly .004 ID
FIX or WORKAROUND
- Unscrew the needle assembly's fitting from injector port 4.
- Unscrew the Seat Port Assembly´s fitting from injector port 1.
- Screw the Seat Port Assembly's fitting in injector port 4.
- Back-flush with water to remove clog
- If back-flushing does not remove clog, replace Seat Port Assembly .004 ID (part number 700005234)
- Return instrument to original configuration