Empower sample set does not move to the next line or injection - WKB87792
Article number: 87792
SYMPTOMS
- A running Empower sample set does not move to the next injection or line in the sample set
- The bottom of the sample set editor window continues to show the current injection still running, even after the specified run time has passed
- The only way to move the sample set to the next injection or line is to click on "Abort" and select "Abort and move to the next injection"
- The project has sufficient table space
- In the run samples window "injection waiting" message may appear
ENVIRONMENT
- Empower 3
- ACQUITY UPLC
- Waters LC
CAUSE
Corrupted system, instrument method, method set, project or communication issue
FIX or WORKAROUND
- Reboot computer
- Create new system WKB2121
- Create a new Empower project.
- To ensure that corruption is not copied to the new project, do not copy or clone the current project or methods.
- If issue remains, create new instrument methods, method sets, and sample set methods.
- Create new instrument methods with only the components needed for the analysis.