Cannot create new system after removing the old system from Empower - WKB61621
Article number: 61621
SYMPTOMS
- Cannot create new system after removing the old system from Empower
- When trying to recreate the system, Empower does not identify any of the instruments configured to LAC/E32
- The ACQUITY system is configured physically to the LAC/E32
- The Instrument tab in node properties is blank, while DHCP lists the IP addresses of the system
- In the properties of the instrument LAN, the card is not configured correctly
ENVIRONMENT
- Empower 3 Enterprise
- ACQUITY UPLC
- Ethernet instruments
CAUSE
The instrument LAN is not configured, causing a communication issue between the LAC/E32 and the ACQUITY system.
FIX or WORKAROUND
- Configure the NIC card to the appropriate address.
- Power-off the system and remove the entries from DHCP.
- Configure the Waters DHCP server.
- Reboot the LAC/E32.
- Open DHCP via C:\Empower\Instruments\Waters DHCP Server Configuration.exe.
- Power-on each system module.
- Confirm that DHCP auto-populates with an IP address for each module.
- Log in to Empower and go to node properties. Select the Instruments tab and scan for instruments. Confirm that all modules appear and show OK? yes.
- Create the new system.
ADDITIONAL INFORMATION
id61621, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPBINARY