Review fails when opening data from a specific project - WKB294063
Article number: 294063
SYMPTOMS
- Review fails when cycling through injections
- Occurs with a specific processing method
- Message center may show "Not enough storage" and "Chrom map view failed" errors
- Loading the method set that contains the problem processing method does not load the PM but loads a blank PM
- Attempts to open the processing method in the project or review gives Database Error; unable to continue


ENVIRONMENT
- Empower
- Review
- Processing Method
CAUSE
The processing method is corrupted.
FIX or WORKAROUND
- Create a new processing method from New.
- If data was previously processed using the problem method:
- Open a results set created using the processing method in Review.
- "Save as" the processing method.
- Choose to overwrite the same name.

- This may allow the processing method to be opened and used again.