Unable to open Empower processing method when versions exceed 1000 - WKB198081

Article number: 198081

SYMPTOMS

  • Database error; unable to continue when attempting to open processing method (PM)
  • Unable to open PM

 clipboard_e209c2f85b200c0167867d5188e86368d.png

ENVIRONMENT

  • Empower
  • FR5 
  • SR5
  • Processing method
  • Versions exceed 1000

CAUSE

Unknown.

The customer was reviewing channels with a complex PM. They saved many times while reviewing a number of channels. If the memory usage of review was high (>2vGb) and the PM version exceeded 1000, this issue could occur.

FIX or WORKAROUND

  1. Do not exceed 1000 versions of a method.
  2. Create a new method at or before version 1000.

ADDITIONAL INFORMATION

Raised as [CRI-3374] 