Removing the ESI probe tip and gasket
Article number: 75863
Required materials
- Chemical-resistant, powder-free gloves
- 7-mm open-end wrench
- 10-mm open-end wrench
To remove the ESI probe tip and gasket:
Warning: To avoid personal contamination with biologically hazardous materials, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid burn injuries, take great care while performing this procedure.
Warning: To avoid puncture injuries, handle sample needles, syringes, fused silica lines, and borosilicate tips with extreme care.
- Remove the ESI probe adapter from the source (see Removing the probe adapter).
- Use the 7-mm wrench in conjunction with the 10-mm wrench to remove the probe tip.
- Remove the metal gasket from the probe tip.
Warning: To avoid spreading contamination with biologically hazardous, toxic, and corrosive materials, dispose of all waste materials according to local environmental regulations.
- Dispose of the metal gasket in accordance with local environmental regulations.
- If the probe tip is damaged, dispose of it in accordance with local environmental regulations.