Skip to main content
Waters

Installing the low-volume vials

Article number: 70277

To install low-volume vials:

1. If a standard reservoir bottle is fitted, remove it.

2. Screw the low-volume adapters into the manifold and finger-tighten them.

 

SELECT SERIES Cyclic IMS Installing Low-Volume Vials.svg

clipboard_e7c6fa0d9d39dc0e7dfaa9522501c5191.png

3. Screw the low-volume vials into the adapters.

4. For each low-volume vial, ensure that the ends of the solvent delivery tubes are positioned so that they are close to, but not touching, the bottom of the vial (see Adjusting the solvent delivery tube positions).

Not able to find a solution? Click here to request help.