Installing the Tool-Free ESI probe adapter
Article number: 75006
Probe adapter parts
Required materials
- Chemical-resistant, powder-free gloves
To install the probe adapter:
Warning: To avoid personal contamination with biologically, hazardous, or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid puncture wounds, handle sharp parts and materials with care.
- Remove the protective cap, if fitted, from the probe tip.
- Carefully slide the probe adapter into the hole in the source enclosure's probe adjuster assembly, ensuring that the probe location dowel aligns with the location hole in the probe adjuster assembly.
- Rotate the probe adapter locking ring clockwise to secure the probe adapter in place.
- Connect the ESI probe adapter’s cable to the high-voltage connector.
- Install the probe assembly. See Installing the probe assembly.