What are the leak test criteria when testing the ccBSM on a UPC2 System? - WKB12978
Article number: 12978
ENVIRONMENT
- ACQUITY UPC2 System
- ACQUITY UPC2 Binary Solvent Manager (ccBSM)
ANSWER
For ccBSM Dynamic Leak Test: < 500 nL/min
- The dynamic leak diagnostic is to be performed on both Pump A (CO2) and Pump B (co-solvent) of the UPC2 ccBSM.
- Running the dynamic leak test on the system side is not recommended.