P-Series pump leaking CO2 - WKB13350

Article number: 13350

SYMPTOMS

  • CO2 leaking from pump heads 
  • Hissing noise because of CO2 leak

ENVIRONMENT

  • P-Series pumps
  • SFX

CAUSE

Bad pump seals

Bad piston

possible bad pump heads

FIX or WORKAROUND

  1. Rebuild the pump heads. 
    1. P-50 seals ( 700007006 ) and backing ( 700006757 )
    2. P-200 seals  ( 700006543 ) and backing ( 700006542 )
    3. P-350 seals   (700007099 ) and backing (  700006813 )
  2. Replace the pistons.
    1. P-200 piston and bearing kit ( 700011415 )
    2. p-350 piston and bearing kit ( 700011416 )

ADDITIONAL INFORMATION

Replace pump heads if none above worked.