SFC150MGM error: Timeout waiting to receive injector start signal - WKB125332

Article number: 125332

SYMPTOMS

"Timeout waiting to receive injector start signal" error reported on SFC150MGM, causing the sample list to stop.

ENVIRONMENT

  • Prep SFC 150 Mgm
  • 2998 Photodiode Array Detector
  • ChromScope 2.1

CAUSE

Total run time of sample list too long, resulting in too much data.

FIX or WORKAROUND

  1. Split the single long run time sample list into multiple sample lists with shorter run time.

    For example: Split a 1200 min sample list into 4 × 300 min sample list.

ADDITIONAL INFORMATION

id125332, MGPREPSFC, SFC150 MGM