How to improve peak shape using passivation with ACQUITY UPLC HSS C18 columns - WKB31187
Article number: 31187
SYMPTOMS
- Peak shape for some alkaloid-type compounds can be suboptimal using ACQUITY UPLC HSS C-18 columns
ENVIRONMENT
- ACQUITY UPLC HSS C-18 columns
- Alkaloid-type compounds such as chloroquine, hydroxychloroquine, and pheniramine
CAUSE
Unclean injector (caused by repeated analysis of these types of compounds)
FIX or WORKAROUND
Employ a passivation technique:
- Disconnect the column from the mass spectrometer, leaving only the LC flow input.
- Make three 10-µL injections of 2% phosphoric acid (H3PO4). This is the passivation step.
- Make ten 10-µL injections of blank mobile phase to flush out the H3PO4.
- (Optional) Store the column in initial mobile phase for three days.
- Reconnect the column to the mass spectrometer and analyze samples.
ADDITIONAL INFORMATION
NOTE: This instruction is for INJECTIONS only. DO NOT USE 2% Phosphoric acid as mobile phase.
For additional information see Can I flush my column with 2 percent Phosphoric acid to passivate the column?
id31187, eluent, UPBINARY