Can I flush with 2% phosphoric acid to passivate the column? - WKB222774
Article number: 222774
ENVIRONMENT
- Column care and use
- ACQUITY UPLC
- HPLC
ANSWER
No, this is not recommended.
ADDITIONAL INFORMATION
2% phosphoric acid is a strong oxidizing acid that when flowed through the mobile phase lines and into a column will clean out the lines and send debris and fouling into the column.
Metal mobile phase filters and sinkers exposed to 2% phosphoric acid will leach metal downstream into the column, shortening its lifetime.
Disconnect the column and remove it from the flow path during a system cleaning or passivation procedure.