Overpressure error on ACQUITY Xevo system - WKB99834

Article number: 99834

SYMPTOMS

  • Overpressure error when flushing system
  • Overpressure disappears when fluidics changed from LC to waste

ENVIRONMENT

  • SQD2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S cronos
  • Xevo TQ-S
  • Xevo TQ-S micro
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

ESI probe clogged.

APCI probe in clogged.

FIX or WORKAROUND

Replace the APCI capillary.
Replace ESI probe / capillary.

ADDITIONAL INFORMATION

 