Overpressure error on ACQUITY Xevo system - WKB99834
Article number: 99834
SYMPTOMS
- Overpressure error when flushing system
- Overpressure disappears when fluidics changed from LC to waste
ENVIRONMENT
- SQD2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S cronos
- Xevo TQ-S
- Xevo TQ-S micro
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS QTof
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
CAUSE
ESI probe clogged.
APCI probe in clogged.
FIX or WORKAROUND
Replace the APCI capillary.
Replace ESI probe / capillary.
ADDITIONAL INFORMATION
id99834, Electrospray, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD