Skip to main content
Waters

How to test for basic sensitivity on the AutoSpec - WKB95113

Article number: 95113

OBJECTIVE or GOAL

To test the basic sensitivity of the AutoSpec to help aid diagnosis of GCMS dioxin sensitivity problems.

ENVIRONMENT

  • AutoSpec Premier
  • AutoSpec Ultima

PROCEDURE

  1. Open all slits.
  2. Use the following instrument conditions (typical dioxin source tune values):
    • Electron Energy = 33 to 35eV
    • Trap current = 650uA
    • Detector voltage = 180V
  3. Empty reference tank, and then inject 2.0 ul of PFK and optimize ion intensity (leave syringe in place).
  4. Ensure that m/z 69Da is >10V signal intensity. Sometimes the barrel of the PFK syringe can be removed from the septum tank to allow air to normalize flow.
  5. Tune instrument to 10,000 resolution to @5% transmission (see WKB95120 for definition and WKB95218 on tuning to 10,000 resolution).
  6. Sensitivity problem is due to the GC/autosampler/standard/sample.

ADDITIONAL INFORMATION

1. If >10V signal is not obtainable on m/z 69, confirm <1000 resolution by closing the collector slit (see WKB94413). If resolution is >1000, source or alpha slit is not correctly set. Contact Waters Service Support.

2. Confirm that the detector gain is within acceptable ranges (see WKB11558) Contact Waters Service Support.

Not able to find a solution? Click here to request help.