How do I correctly position the collector slit during tuning on the AutoSpec? - WKB94413
Article number: 94413
OBJECTIVE or GOAL
To correctly position the collector slit position when tuning the AutoSpec.
ENVIRONMENT
- AutoSpec Premier (P)
- AutoSpec Ultima (M)
PROCEDURE
- Obtain a square-topped peak and adjust Peak Display so that the peak top is at 100%.

- Close collector slit without reducing the intensity below the 75% threshold line.

ADDITIONAL INFORMATION
1. Over-closing the collector slit can adversely affect the instrument's sensitivity (particularly at resolutions >5000).
2. Can mask other AutoSpec tuning problems.