QDa status light continually flashes orange; no communication with MassLynx - WKB9511
Article number: 9511
SYMPTOMS
- Lost power to the system
- Powered-on the QDa, the PC, and then the ACQUITY I-Class components
ENVIRONMENT
- QDa
- ACQUITY I-Class system
- MassLynx 4.1
CAUSE
The QDa was powered-on before the PC and ACQUITY components.
FIX or WORKAROUND
- Power-on the PC and allow it to fully initialize.
- Power-on the ACQUITY components.
- Power-on the QDa.
- Allow all instruments to finish initializing, and then open MassLynx.