TQ-S micro vents following cone change - WKB85409
Article number: 85409
SYMPTOMS
- Instrument vents after cone change
- Different cone, nozzle, and O-ring do not resolve issue
ENVIRONMENT
- All Xevo mass spectrometers
CAUSE
The M3 socket shoulder screw that secures the upper gas cone position block became loose, causing a source leak between cone and O-ring.
FIX or WORKAROUND
- Tighten the M3 socket shoulder screw that holds on the gas cone position block.
- The shoulder screw and/or Belleville washers may need to be replaced.
ADDITIONAL INFORMATION
700004404 - M3 shoulder screw (2 pk)
700003080 - Belleville spring washer (10 pk)
id85409,