SYMPTOMS
- The instrument was recently pumped down, and the message "Waiting for TOF Pressure" appears
- TOF Vacuum is at 1.28e-6 mbar
ENVIRONMENT
- Xevo G2-XS
CAUSE
The instrument has not yet reached operating pressure.
The instrument needs to be at 1.2e-6 mbar or below for it to enter Operate mode.
FIX or WORKAROUND
Wait for the ToF pressure to fall below 1.2e-6 mbar.
ADDITIONAL INFORMATION
id85480, XEVOG2XSTF