MS scan detects the spectrum of the target compound cleavage; caused by collision energy - WKB74705
Article number: 74705
SYMPTOMS
- MS scan detects the spectrum of the target compound cleavage
ENVIRONMENT
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo G2-S
- Xevo G2-XS
CAUSE
It collides with the argon gas in the collision cell and decays.
FIX or WORKAROUND
When Collision Energy in the MS Method is set to Off, a 6V collision voltage is applied.
As shown in the figure below, after setting Collision Energy to On, specify a value smaller than 6V (e.g., 3V).
If the problem persists, see the following article.
MS scan detects the spectrum of the target compound cleavage: Caused by temperature
MS scan detects the spectrum of the target compound cleavage: Caused by StepWave
ADDITIONAL INFORMATION
The following ions tend to undergo in-source decay.
– Ions less than m/z 300
– Ions with high valence
– Ions with unstable structure