MS scan detects the spectrum of the target compound cleavage: Caused by temperature - WKB74701
Article number: 74701
SYMPTOMS
MS scan detects the spectrum of the target compound cleavage.
ENVIRONMENT
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo G2-S
- Xevo G2-XS
CAUSE
High desolvation temperature.
FIX or WORKAROUND
- Open the Tune page.
- Select the mode (Positive/Negative, Sensitivity/Resolution) to be used for analysis.
- Lower the Desolvation in the Temparature column to 300° C.
If the problem persists, see the following articles.
MS scan detects the spectrum of the target compound cleavage: Caused by StepWave
MS scan detects the spectrum of the target compound cleavage: Caused by Collision Energy
ADDITIONAL INFORMATION
The following ions tend to undergo in-source decay.
– Ions less than m/z 300
– Ions with high valence
– Ions with unstable structure
id74701, MALDIG2SI, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF