Sensitivity low on the Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro or Xevo TQ-XS - WKB72231
Article number: 72231
SYMPTOMS
- Lower than expected sensitivity
- While infusing a sample and monitoring the beam on the tune page, gently pushing or pulling the probe results in beam intensity increase
- System is not charging and all voltages are as expected
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Probe tip is slightly bent and not directing the nebulized spray toward the sample cone.
FIX or WORKAROUND
Replace or straighten the probe tip.
ADDITIONAL INFORMATION