Xevo TQ-XS sensitivity low - WKB72231

Article number: 72231

SYMPTOMS

  • Lower than expected sensitivity
  • While infusing a sample and monitoring the beam on the tune page, gently pushing or pulling the probe results in beam intensity increase
  • System is not charging and all voltages are as expected

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Probe tip is slightly bent and not directing the nebulized spray toward the sample cone.

FIX or WORKAROUND

Replace or straighten the probe tip.

ADDITIONAL INFORMATION

 