Sensitivity low on the Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro or Xevo TQ-XS - WKB72231
Article number: 72231
SYMPTOMS
- Lower than expected sensitivity
- While infusing a sample and monitoring the beam on the tune page, gently pushing or pulling the probe results in beam intensity increase
- System is not charging and all voltages are as expected
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Probe tip is slightly bent and not directing the nebulized spray toward the sample cone.
FIX or WORKAROUND
Replace or straighten the probe tip.
ADDITIONAL INFORMATION
id72231, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD