Skip to main content
Waters

Sensitivity low on the Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro or Xevo TQ-XS - WKB72231

Article number: 72231

SYMPTOMS

  • Lower than expected sensitivity
  • While infusing a sample and monitoring the beam on the tune page, gently pushing or pulling the probe results in beam intensity increase
  • System is not charging and all voltages are as expected

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Probe tip is slightly bent and not directing the nebulized spray toward the sample cone.

FIX or WORKAROUND

Replace or straighten the probe tip.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.