Solvent leaks from the fitting for the tube that runs from the diverter valve to the probe on a Xevo TQD - WKB72264
Article number: 72264
SYMPTOMS
- Leak at the fitting for the tube that runs from the diverter valve to the probe
- Tightening the fitting does not resolve
- Replacing the PEEK fitting does not resolve
ENVIRONMENT
- Xevo TQD
CAUSE
Damaged PEEK tubing
FIX or WORKAROUND
- Remove the tubing and the fitting from the valve.
- Cut approximately 1/2 inch from the end of the PEEK tubing.
- Reconnect the tubing and the fitting to the valve.
ADDITIONAL INFORMATION