High RSD on Xevo TQ-XS in combination with APGC - WKB68661
Article number: 68661
SYMPTOMS
- When running a long sequence of the same sample (more than 20 injections), the RSD is over 10%
ENVIRONMENT
- Xevo TQ-XS
- APGC
- CTC PAL
CAUSE
When tuning on the column bleeding peak (m/z 207), the beam intensity is unstable.
The CTC PAL autosampler syringe does not enter the middle of the injection port septa.
FIX or WORKAROUND
- Adjust the Aux gas flow until the bleeding peak is stable.
- Check the injector parameters of the CTC and adjust the position is needed.
ADDITIONAL INFORMATION