What type of power receptacle is required for a PowerVar ABCDEF5200-22UPS? - WKB61012
Article number: 61012
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC H-Class PLUS
- SQ Detector 2
- Xevo TQD
ANSWER
In North America, the PowerVar Security Plus UPS 5.2 kVA requires a NEMA L6-30 receptacle rated at 25 A.
ADDITIONAL INFORMATION
On the UPS, there are three duplex NEMA 5-20R receptacles (six receptacles total) and three L6-15R receptacles.
For specifications, follow the link and select the ABCDEF5200-22 model: https://www.powervar.com/products/si...lus-series-ups
The ABCDEF5200-22 is marketed only in North America as a regional variety of the worldwide model ABCDEF6000-22 with modified firmware to ensure that it will not overload a 208-V supply rated 25A .