Reduced peak sensitivity for drug material standards on SQD - WKB60930

Article number: 60930

SYMPTOMS

  • Peak areas of standard peaks reduce or drop in intensity over many assays
  • Cleaning cone partially brings back signal

ENVIRONMENT

  • ACQUITY SQD
  • ACQUITY TQD
  • Xevo TQD
  • Xevo TQMS
  • Xevo TQS

CAUSE

Dirty ESI probe capillary

FIX or WORKAROUND

Rebuild or replace the ESI probe.

ADDITIONAL INFORMATION

 

id60930, A-SQ, A-SQ, Electrospray, SQD, SQD2, TQD, TQDIVD, XEVOTQD, XEVOTQDIVD