How to split LC flow using MS fluidics - WKB60766
Article number: 60766
OBJECTIVE or GOAL
Split LC flow using MS fluidics.
ENVIRONMENT
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo G2-S
- Xevo G2-XS
- Vion IMS QTof
- UNIFI
- MassLynx
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- SQ Detector
- ACQUITY SQD
- ACQUITY RDa Detector
PROCEDURE
- Cut a PEEK tubing to the desired split ratio. For example, to achieve a 1-to-3 split using 60 cm of PEEK tubing, cut the tubing to 20 cm (1/3) and 40 cm (2/3) lengths.
- Disconnect the tubing on port 3 of the fluidics diverter valve.
- Connect the shorter length of tubing to port 3, and place the outlet in a waste bottle.
- Connect the longer length of tubing from port 2 to the ESI probe.
- Configure the events in the acquisition software to combine flow.
ADDITIONAL INFORMATION
Flow to waste and flow to LC still work with the modified configuration.
Can also be used to avoid UV/PDA/FLR flow cell overpressure when the optical detector is connected directly to MS. Naturally, the disconnected channel cannot be used for tuning/calibration/post-column addition.
