Desolvation temperature fails to settle - WKB56412

Article number: 56412

SYMPTOMS

  • Desolvation temperature fails to settle 
  • Desolvation temperature does not rise above 47 °C

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro

CAUSE

Corrupt temporary tune files

FIX or WORKAROUND

  1. Close MassLynx and the MS Console.
  2. Navigate to the C:/MassLynx/Harrier folder. *
  3. In the Harrier folder, find and delete the following temporary tune files: nonstorageexp.exp, tuneexp.exp, and tuneprm.ipr
  4. Launch MassLynx, open the Tune page, and set the desired desolvation temperature.

ADDITIONAL INFORMATION

Temporary tune files are recreated when MassLynx opens.

*The folder is named Harrier for XEVO TQD but have different names depending on MS. Find what it is called for your instrument in WKB62726.