Xevo TQ-S will not go into Operate - WKB5642
Article number: 5642
SYMPTOMS
- Xevo TQ-S will not go into Operate
- Found: TQ-S is vented
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
CAUSE
The mass spectrometer is vented.
FIX or WORKAROUND
On the Tune page, go to Vacuum > Pump.
ADDITIONAL INFORMATION
No Waters mass spectrometers can be put into Operate if the system is vented.