Isolation valve stuck - WKB50539

Article number: 50539

SYMPTOMS

Isolation valve is stuck. Cannot easily move from blocked to open.

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof
  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • ACQUITY SQD system
  • SQ Detector
  • ACQUITY TQD system
  • TQ Detector
  • SQ Detector 2

CAUSE

Isolation valve is warped or damaged.        

FIX or WORKAROUND

  1. Vent and remove isolation valve.
  2. Replace the Perlast O-ring. (Replace the stem also if showing signs of damage.)

ADDITIONAL INFORMATION