No sample peaks in chromatogram with ACQUITY SQD - WKB50547
Article number: 50547
SYMPTOMS
- No peaks appear in chromatogram
- Infusion of setup solution shows very low signal, regardless of tuning parameters
- ESI capillary sticks more than 5 mm out of the probe tip
- Left/right adjustment of probe does not resolve the issue
ENVIRONMENT
- ACQUITY SQD
- ACQUITY TQD
CAUSE
The ESI capillary tip extends too far out of the probe tip.
ESI spray is below the sample cone opening.
FIX or WORKAROUND
Rebuild the ESI probe, ensuring that the steel capillary ends up only 1-2 mm out of the end of the probe tip.
ADDITIONAL INFORMATION
id50547, A-SQ, Electrospray, SQD, TQD, TQDIVD