UniSpray probe leaks - WKB4787
Article number: 4787
SYMPTOMS
- Solvent leaks from the end of the white heat shrink section of the probe (the part with the probe part number written on it)
ENVIRONMENT
- Xevo TQ-XS
- UniSpray source
CAUSE
Incorrectly fitted probe
FIX or WORKAROUND
- Reseat the probe in the diversion valve port, ensuring that there is no dead volume.
- Tighten the PEEK fitting fully. The fitting will "click" when it is fully tight.
ADDITIONAL INFORMATION
id4787, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD