With Vion, lockmass error after refill, but beam is stable once recovered - WKB47792
Article number: 47792
SYMPTOMS
- The lockspray takes too long to recover after a refill
ENVIRONMENT
- Vion
CAUSE
The stator face is damaged.
FIX or WORKAROUND
- Order a new HT Selector Valve POD (70000946).
ADDITIONAL INFORMATION
id47792, VION