With Vion, lockmass error after refill, but beam is stable once recovered - WKB47792

Article number: 47792

SYMPTOMS

  • The lockspray takes too long to recover after a refill

ENVIRONMENT

  • Vion

CAUSE

The stator face is damaged.

FIX or WORKAROUND

  1. Order a new HT Selector Valve POD (70000946).

ADDITIONAL INFORMATION

 