"Turbo Controller Fault" on Xevo TQD - WKB47715

Article number: 47715

SYMPTOMS

  • MS vented
  • Turbo Controller Fault error message
  • Low vacuum error message

ENVIRONMENT

  • Xevo TQD
  • SQD2
  • MassLynx

CAUSE

The straight fitting on the NW25 scroll pump tee connecting the 10-mm source backing tubing is loose.

FIX or WORKAROUND

Replace 10-mm straight fitting (700005810) or add a small amount of PTFE tape to fitting to thread to prevent movement. 