Penning gauge needs frequent replacing or cleaning - WKB297819
Article number: 297819
SYMPTOMS
- Analyzer MS2 Penning gauge reads inaccurately (for example, 1.0 e-8 mbar or 1.0 e-2 mbar)
- Failure reoccurs within a few months of replacing the Penning gauge body
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQ Absolute
CAUSE
The vacuum window gauge is left open on the tune page. This switches the Penning gauge on at all times, which leads it to become dirty quickly.
FIX or WORKAROUND
Open the vacuum gauge window on the tune page only when you need to monitor pressures within the instrument.