No spray from tool-free ESI probe used on SQD2, Xevo G2-S, Xevo TQ Absolute, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro, or Xevo TQ-XS - WKB278387
Article number: 278387
SYMPTOMS
- No spray from ESI tool-free probe
- No signal observed on Tune page
- Diverter valve is set to LC
- UPLC system pressure is slightly higher than normal
- Disconnecting the probe assembly from the diverter valve shows solvent reaching the diverter valve
ENVIRONMENT
- Xevo TQ-S micro
- Xevo TQD
- Xevo TQ-XS
- Xevo G2-S
- Xevo TQ Absolute
- SQD2
- Xevo TQ-S
CAUSE
The ESI assembly for the tool-free probe is blocked or damaged.
FIX or WORKAROUND
Replace the ESI assembly for the tool-free probe. See KB article WKB85646 for part numbers.
ADDITIONAL INFORMATION
id278387, A-SQ, Electrospray, SQD2, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSQT, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD