Skip to main content
Waters

No spray from tool-free ESI probe used on SQD2, Xevo G2-S, Xevo TQ Absolute, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro, or Xevo TQ-XS - WKB278387

Article number: 278387

SYMPTOMS

  • No spray from ESI tool-free probe
  • No signal observed on Tune page 
  • Diverter valve is set to LC
  • UPLC system pressure is slightly higher than normal
  • Disconnecting the probe assembly from the diverter valve shows solvent reaching the diverter valve

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-XS
  • Xevo G2-S
  • Xevo TQ Absolute
  • SQD2
  • Xevo TQ-S

CAUSE

The ESI assembly for the tool-free probe is blocked or damaged.

FIX or WORKAROUND

Replace the ESI assembly for the tool-free probe. See KB article WKB85646 for part numbers. 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.