Low sensitivity on the SQD2, Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro and Xevo TQ-XS - WKB244245
Article number: 244245
SYMPTOMS
- Peaks of interest have lower sensitivity or are missing
- Cleaning cone does not resolve issue
- All LC diagnostics pass
ENVIRONMENT
- MassLynx
- SQD2
- Xevo TQ-MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ-S micro
CAUSE
The wrong calibration is loaded on the tune page.
FIX or WORKAROUND
Link the newest tune calibration to the current tune page.