Lockmass Fluidics Check error on the BioAccord system - WKB244202
Article number: 244202
SYMPTOMS
- After running out of lockmass solution during an analysis, bottle was replaced
- System gives Lockmass Fluidics Check health status
- Selecting the Re-check feature results in the same failed status
ENVIRONMENT
- waters_connect 2.1.2
- BioAccord System
CAUSE
Incorrectly made lockmass solution; no lockmass concentrate added to the diluent bottle
FIX or WORKAROUND
- Correctly make lockmass solution of part 186009298 by adding the concentrate to the diluent bottle.
- Install newly made solution on the ACQUITY RDa Detector.
- On the Instrument Summary page, select the bar for "Click here to re-check system" for the test to rerun.