Spray from ESI probe is unstable, signal comes in bursts, on the tool-free ESI probe using the SQD2, Xevo G2-S, Xevo TQ Absolute, Xevo TQD, Xevo TQ-S, Xevo TQ-S micro or Xevo TQ-XS - WKB241285

Article number: 241285

SYMPTOMS

  • Spray is unstable
  • Signal comes in bursts

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQD
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo G2-S
  • Xevo TQ Absolute
  • SQD2
  • Tool-free ESI probe

CAUSE

Partially blocked probe assembly

FIX or WORKAROUND

Replace the ESI probe assembly. See KCS Article WKB85646 for specific part numbers.

ADDITIONAL INFORMATION

 

id241285, A-SQ, Electrospray, SQD2, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSQT, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.