SYNAPT G2-S message "Fluidics Error, Select Purge to recover" will not clear - WKB24097
Article number: 24097
SYMPTOMS
- "Fluidics Error, Select Purge to recover"
- Selecting Purge just brings the error back
ENVIRONMENT
- SYNAPT G2-S
- MassLynx
CAUSE
There is an error in the background. MassLynx and the console can report only one error at a time.
FIX or WORKAROUND
- Inspect the console log to see if the original error is listed.
- Take the appropriate action to resolve that error.
- Purge the fluidics.
ADDITIONAL INFORMATION
If there is no obvious error, try opening and closing the source door, wait a minute or two, and then purge the fluidics.