Source Exhaust Pressure trip with Xevo instrument - WKB240851

Article number: 240851

SYMPTOMS

  • Source Exhaust Pressure trip

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-XS

CAUSE

Leak from tubings on the source exhaust vent line    

FIX or WORKAROUND

  1. Remove and reseat the tubing connectors from the source exhaust vent on the MS to the trap bottle.
  2. Trim tubing if required to create a new area to fit into the push-fit fitting.

ADDITIONAL INFORMATION